Standard Event Status Register
Table 5 Standard Event Status Register Bits
Bit Description
0 Operation Complete. A 1 in this bit position indicates that all pending N8211A/N8212A operations
were completed following execution of the *OPC command.
1 Request Control. This bit is always set to 0. (The N8211A/N8212A does not request control.)
2 Query Error. A 1 in this bit position indicates that a query error has occurred. Query errors have
instrument error numbers from −499 to −400.
3 Device Dependent Error. A 1 in this bit position indicates that a device dependent error has occurred.
Device dependent errors have instrument error numbers from −399 to −300 and 1 to 32767.
4 Execution Error. A 1 in this bit position indicates that an execution error has occurred. Execution errors
have instrument error numbers from −299 to −200.
5 Command Error. A 1 in this bit position indicates that a command error has occurred. Command errors
have instrument error numbers from −199 to −100.