User`s guide
Operating the DC Power Analyzer 3
Model N6705A User’s Guide 69
Standard (interleaved)
Standard (interleaved) mode only applies when both voltage and
current measurement traces are selected on power modules other
than Agilent N676xA. These other power modules cannot measure
voltage and current simultaneously; hence, the voltage and current
measurements must be interleaved. Each measurement is sampled
for about 5 milliseconds at the beginning of every sample period.
Power is calculated from the interleaved measurements. Standard
data sampling is used for the following Module/Trace selections:
Power Module Display Trace selection Available capabilities set for ALL outputs
N673xB, N674xB Voltage and Current, or Power ● Sample period = 75 ms to 60 s
N675xA, N677xA Voltage and Current, or Power ● Trigger source = Run/Stop key only
● Trigger offset = 0 (offset not available)
● Values logged = average only
Sample period
(Min. = 1 ms; Max. = 60 s)
Sample period
Data point
Sample rate = 50 kHz
Average
value (V or I)
Min. value
(V or I)
Max. value
(V or I)
Voltage measurement
250 samples @ 50 kHz
Current measurement
250 samples @ 50 kHz
Sample period
(Min. = 75 ms; Max. = 60 s)
wait
Sample period
wait
Average
value (V)
Average
value (I)