User`s guide
3 Operating the DC Power Analyzer
58 Model N6705A User’s Guide
Using the Waveform Display Knobs
Knob: Description:
Vertical Volts/Div
Makes the waveform bigger or smaller vertically in relation to its
ground reference. Specified in volts/division or amps/division on the
y axis. If the vertical gain causes the trace to be out of view, arrow
symbols will indicate the direction of the trace.
Vertical Offset
Moves the ground reference of the trace up or down in relation to the
horizontal center line of the grid. The offset popup that
appears in the upper right corner of the display shows
how far the ground reference of the selected trace is above or below
the horizontal center line of the grid.
Positive values indicate how far the center line is above the ground
reference. Negative values indicate how far the center line is below
the ground reference.
Horizontal Time/Div
Stretches or shrinks the waveform horizontally around the horizontal
offset reference. Specified in time/division on the x axis. The time-
base applies to ALL output traces.
Horizontal Offset
Moves the waveform to the right or left of the horizontal offset
reference. The trigger point of the waveform is indicated by the solid
arrow.
Trigger Level
Moves the trigger level up and down when a voltage or current level
is the trigger source. The trigger level is identified by the symbol.
If the trigger level is out of view, an arrow symbol
will indicate
the direction of the trigger level.
Marker 1/Marker 2
Moves the measurement markers right or left on the display. Press
Scope View to display the markers. Values at the bottom of the
display are referenced to the intersection of the markers. If a marker is
out of view, an arrow symbol indicates the direction of the marker.
Push the Market1/Marker2 knobs to reset the markers.