User`s guide

3 Operating the DC Power Analyzer
56 Model N6705A User’s Guide
Symbol/Field: Description:
5 Scope/Arb Status
Indicates whether the scope is Running, Stopped, or Waiting for a
trigger.
6 Trigger Mode
Sample Rate
Indicates the trigger mode (Auto, Single, or Triggered).
The indicated scope sample rate is based on the horizontal time/div.
setting. When the time/div. setting is less than 20 ms/division, the
scope will sample at its fastest rate: 20.48 microseconds.
7 Output
Pop-up
If you turn the voltage and current knobs,
a
pop-up dialog will indicate the present
output settings.
8 Out of View Arrows
Indicates that the trace, V1 in this example, is out of view. Use the
Vertical Volt/Div knob or the Vertical Offset knob to bring the trace
into view.
Push the Trigger Level knob to autoscale the traces so that all of them
will appear on the display.
9 Ground Reference
The ground reference of the trace. Ground references are offset so
that they do not overlap. The ground reference offset value is
referenced to the horizontal center line of the grid.
10 Power Traces
Labels for the power traces appear on the center of the grid (P1, P2,
P3, P4). Traces are color coded according to output. Note that only
Agilent Models N6761A/N6762A. can display power traces.
Push the Trigger Level knob to autoscale the traces.
11 Trigger Level
Shows the location of the voltage or current trigger level and output.
In this example, the voltage trigger level of output 2 is shown. The
trigger source and amplitude are shown at the bottom right of the
display.
12 Current Trace
Labels for the current traces appear on the right side of the grid (I1, I2,
I3, I4). Traces are color coded according to output.
Push the Trigger Level knob to autoscale the traces so that all of them
will appear on the display.
13 Trigger Source
Amplitude
The scope’s trigger source. In this example, the trigger source is a
voltage level on output 2.
Indicates the measurement will be triggered
on the up-slope (positive).
Indicates the measurement will be triggered
on the down-slope (negative).
If the trigger source is set to a voltage or current level, the amplitude
of the trigger level is indicated below the trigger source. In this
example, the voltage trigger level is set to 4.5V.
14 Time/Div.
Trigger Offset
Identifies the horizontal time-base setting. This can be adjusted using
the front panel Horizontal Time/Div knob.
The trigger offset indicates the time from the trigger point to the
offset reference. Use the front panel Horizontal Offset knob to adjust.