Operating the DC Power Analyzer 3
Model N6705A User’s Guide 53
Trigger the Arb
NOTE
For the arbitrary waveform to appear at the output terminals, the selected
output must be turned on before the Arb is run.
Depending upon the selected trigger source, you can trigger the
arbitrary waveforms as follows:
Trigger Source: Description:
Arb Run/Stop key
Press the Arb Run/Stop key to start the
Arb waveform. Press the key again to stop
the Arb waveform.
Rear Trigger input
Provide a low-true signal to the rear
trigger input BNC connector. The signal
must persist for at least 10 milliseconds.
Remote command
Send a remote trigger command over one
of the three interfaces (i.e. *TRG).
Once configured, the instrument will wait indefinitely for the trigger
signal. If the trigger does not occur, and you wish to cancel the
arbitrary waveform, press the Arb Run/Stop key to stop the Arb.
After a trigger is received and the arbitrary waveform completes, the
Voltage After Arb setting determines what the output does. If the
Return to DC Value box is checked, the output voltage and current
return to the settings that were in effect before the arbitrary
waveform started. If the Last Arb Value box is checked, the output
will remain at the last Arb setting.