User`s guide

3 Operating the DC Power Analyzer
48 Model N6705A User’s Guide
Sine Properties
The following window programs the Sine properties:
Parameter: Description:
Amplitude (V
0
)
Frequency (f)
Offset (V
1
)
The amplitude or peak value.
The frequency of the sine wave.
The offset from zero.
Because the output cannot generate
negative voltages, the offset cannot be
less than the amplitude.
Pulse Properties
The following window programs the Pulse properties:
Parameter: Description:
Start Voltage (V
0
)
Pulse Voltage (V
1
)
Delay (T
0
)
Pulse Width (T
1
)
End Time (T
2
)
The voltage before and after the pulse.
The voltage of the pulse.
The delay after the trigger is received.
The width of the pulse.
The time V0 persists after the pulse.