User`s guide

2 Installation
26 Model N6705A User’s Guide
Power Module Location
Detailed information about installing and removing the power
modules is provided in the Agilent N6705A Service Guide. It is
recommended that this be done by qualified service personnel.
The location of the power modules inside the mainframe determines
to which front panel output terminals they are connected. To view
the power module/output terminal assignments, turn the unit on,
press the Settings key, then press Properties. The power modules are
listed under each output channel.
Outputs that are not connected to a power module will not be
displayed in the Meter view.
Bench Installation
Do not block the air intake and exhaust at the sides, or the exhaust
at the rear of the unit. Refer to the outline diagram in Appendix A.
Minimum clearances for bench operation are 2 inches (51 mm) along
the sides and back.
For easier display viewing and binding post access, you can tilt the
front of the unit up by rotating the extension bar down.
Rack Installation
CAUTION
Use Rack Mount kit (Option 908 or Option 909 with handles) to rack mount the
instrument.
Agilent N6705A DC Power Analyzer mainframes can be mounted in a
19-inch EIA rack cabinet. They are designed to fit in four rack-units
(4U) of space.
Remove the feet before rack mounting the unit. Do not block the air
intake and exhaust at the sides of the unit, or the exhaust at the rear
of the unit.
Cleaning
WARNING
SHOCK HAZARD To prevent electric shock, unplug the unit before cleaning.
Use a dry cloth or one slightly dampened with water to clean the
external case parts. Do not use detergent or chemical solvents. Do
not attempt to clean internally.