User`s guide
Appendix D Output On/Off Synchronization
144 Model N6705A User’s Guide
Digital Connections and Configuration
NOTE
Only pins 4 through 7 can be configured as synchronization pins. You cannot
configure more than one On Couple and one Off Couple pin per mainframe.
The polarity of the pins is not programmable; it is set to Negative.
The digital connector pins of the synchronized mainframes that
contain coupled outputs must be connected together as shown in the
following figure. In this example, pin 6 will be configured as the
output On control. Pin 7 will be configured as the output Off control.
The ground or Common pins also need to be connected together.
Only two of the digital connector pins on each mainframe can be
configured as “On Couple” and “Off Couple” on each synchronized
mainframe. The designated pins will function as both an input and an
output, with a negative transition on one pin providing the
synchronization signal to the other pins.
Operation
Once configured and enabled, turning the output on or off on any
coupled output will cause all coupled outputs on all configured
mainframes to turn on or off according to their user-programmed
delays. This applies to the front panel On/Off keys, the Web server,
and to SCPI commands.
Turning the outputs on or off using the front panel All Outputs
On/Off keys will cause all coupled outputs as well as non-coupled
outputs on that mainframe to turn on or off.
1 2 3 4 5 6 7
I
1 2 3 4 5 6 7
I
1 2 3 4 5 6 7
I
N6705A #1 N6705A #2 N6705A #3
On Couple
Off Couple
I