User`s guide
SCPI Commands Appendix C
Model N6705A User’s Guide 139
SCPI Command Description
TRIGger
:ACQuire (ACQuire commands only on N6761A/62A and Opt. 054)
[:IMMediate] (@chanlist) Triggers the measurement immediately
:SOURce BUS|PIN<1-7>|TRANsient<1-4>, (@chanlist) Sets the measurement trigger source
:ARB
:SOURce IMMediate | EXTernal | BUS Sets the ARB trigger source (only on N6705A)
:DLOG (DLOG commands only on N6705A)
[:IMMediate] Triggers the data logger immediately
:CURRent
[:LEVel] <NRf>, (@chanlist) Sets the current trigger level of the data logger
:SLOPe POSitive | NEGative, (@chanlist) Sets the current trigger slope of the data logger
:SOURce IMMediate | EXTernal | BUS | VOLTage<1-4> Sets the source of data logger trigger
| CURRent<1-4>| ARSKey | OOOKey
:VOLTage
[:LEVel] <NRf>, (@chanlist) Sets the voltage trigger level of the data logger
:SLOPe POSitive | NEGative, (@chanlist) Sets the voltage trigger slope of the data logger
:TRANsient
[:IMMediate] (@chanlist) Triggers the output immediately
:SOURce BUS|PIN<1-7>|TRANsient<1-4>, (@chanlist) Sets the output trigger source
Common Commands
Command Description Command Description
*CLS Clear status *RST Reset
*ESE <NRf> Standard event status enable *SAV <NRf> Saves an instrument state
*ESR? Return event status register *SRE <NRf> Set service request enable register
*IDN? Return instrument identification *STB? Return status byte
*OPC Enable "operation complete" bit in ESR *TRG Trigger
*OPT? Return option number *TST? Performs self-test, then returns result
*RCL <NRf> Recalls a saved instrument state *WAI Pauses additional command processing
*RDT? Return output channel descriptions until all device commands are done