User`s guide

Appendix C SCPI Commands
134 Model N6705A User’s Guide
SCPI Command Summary
NOTE
Some [optional] commands have been included for clarity. All settings commands have a
corresponding query. Not all commands apply to all models.
SCPI Command Description
ABORt
:ACQuire (@chanlist) Resets the measurement trigger system to the Idle state
:DLOG Stops a running data log (only on N6705A)
:TRANsient (@chanlist) Resets the transient trigger system to the Idle state
CALibrate
:CURRent
[:LEVel] <NRf>, (@channel) Calibrates the output current programming
:MEASure <NRf>, (@channel) Calibrates the current measurement
:PEAK (@channel) Calibrates the peak current limit (only on N675xA/N676xA)
:DATA <NRf> Enters the calibration value
:DATE <SPD>, (@channel) Sets the calibration date
:DPRog (@channel) Calibrates the current downprogrammer
:LEVel P1 | P2 | P3 Advances to the next calibration step
:PASSword <NRf> Sets the numeric calibration password
:SAVE Saves the new cal constants in non-volatile memory
:STATE <Bool> [,<NRf>] Enables/disables calibration mode
:VOLTage
[:LEVel] <NRf>, (@channel) Calibrates the output voltage programming
:CMRR (@channel) Calibrates common mode rejection ratio (only N675xA/N676xA)
:MEASure <NRf>, (@channel) Calibrates the voltage measurement
DISPlay
[:WINDow]:VIEW METER1 | METER4 Selects 1-channel or 4-channel meter view
FETCh (FETCh commands only on N6761A/62A and Opt. 054)
[:SCALar]
:CURRent [:DC]? (@chanlist) Returns the average output current
:VOLTage [:DC]? (@chanlist) Returns the average output voltage
:ARRay
:CURRent [:DC]? (@chanlist) Returns the instantaneous output current
:VOLTage [:DC]? (@chanlist) Returns the instantaneous output voltage
HCOPy (HCOPy commands only on Agilent N6705A)
:SDUMp:DATA? Returns an image of the display in .gif format
INITiate
[:IMMediate]
:ACQuire (@chanlist) Enables measurement triggers (only N6761A/62A and Opt. 054)
:DLOG “filename” Enables the Data Logger function (only on N6705A)
:TRANsient (@chanlist) Enables output triggers
:CONTinuous
:TRANsient <Bool>, (@chanlist) Enables/disables continuous transient triggers