User`s guide

Appendix B Error Messages
132 Model N6705A User’s Guide
Selftest Errors (these errors set Standard Event Status register bit #3)
202 Selftest Fail Aux Adc 0 expected <n1> to <n2>, measured <n3>, chan <n4>
Auxiliary ADC failed. n1 and n2 are the expected limits. n3 is the measured value. n4 is the
channel location of the failed module.
202 Selftest Fail DACs 0 expected <n1> to <n2>, measured <n3>, chan <n4>
Both voltage and current DACs are at zero. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.
202 Selftest Fail DACs 1 expected <n1> to <n2>, measured <n3>, chan <n4>
Voltage DAC is at zero; current DAC is at full scale. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.
202 Selftest Fail DACs 2 expected <n1> to <n2>, measured <n3>, chan <n4>
Voltage DAC is at full scale; current DAC is at zero. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.
202 Selftest Fail DACs 3 expected <n1> to <n2>, measured <n3>, chan <n4>
Both voltage and current DACs are at full scale. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.