User`s guide
Error Messages Appendix B
Model N6705A User’s Guide 129
Device-dependent Errors (continued)
207 Cal file version error
The calibration file was written or read using old firmware. Firmware must be updated.
302 Option not installed
The option that is programmed by this command is not installed.
303 There is not a valid acquisition to fetch from
There is no valid data in the measurement buffer.
304 Volt and curr in incompatible transient modes
Voltage and current cannot be in Step and List mode at the same time.
305 A triggered value is on a different range
A triggered value is on a different range than the one that is presently set.
306 Too many list points
Too many list points have been specified.
307 List lengths are not equivalent
One or more lists are not the same length.
308 This setting cannot be changed while transient trigger is initiated
Setting cannot be changed while the instrument is waiting for or executing a trigger sequence.
309 Cannot initiate, voltage and current in fixed mode
Cannot initiate the transient generator because either the voltage or the current function is set
to Fixed mode.
Command Errors (these errors set Standard Event Status register bit #5)
−100
Command error
Generic syntax error.
−101
Invalid character
An invalid character was found in the command string.
−102
Syntax error
Invalid syntax was found in the command string. Check for blank spaces.
−103
Invalid separator
An invalid separator was found in the command string. Check for proper use of , ; :
−104
Data type error
A different data type than the one allowed was found in the command string.
−105
GET not allowed
A group execute trigger is not allowed in a command string.
−108
Parameter not allowed
More parameters were received than were expected.
−109
Missing parameter
Fewer parameters were received than were expected.
−110
Command header error
An error was detected in the header.
−111
Header separator error
A character that was not a valid header separator was found in the command string.
−112
Program mnemonic too long
The header contains more than 12 characters.
−113
Undefined header
A command was received that was not valid for this instrument.
−114
Header suffix out of range
The value of the numeric suffix is not valid.