User`s guide

Appendix B Error Messages
128 Model N6705A User’s Guide
Error List
Error Device-dependent Errors (these errors set Standard Event Status register bit #3)
0 No error
This is the response to the ERR? query when there are no errors.
100 Too many channels
You have specified more channels than are installed in the mainframe.
101 Calibration state is off
Calibration is not enabled. The instrument will not accept calibration commands.
102 Calibration password is incorrect
The calibration password is incorrect.
103 Calibration is inhibited by switch setting
Calibration mode is locked out by the calibration switch.
104 Bad sequence of calibration commands
Calibration commands have not been entered in the proper sequence.
105 Unexpected output current
The measured output current is outside the acceptable range.
106 Zero measurement out of range error
The “zero” measurement value is outside the acceptable range.
107 Programming cal constants out of range
The programmed calibration constant is outside the acceptable range.
108 Measurement cal constants out of range
The measurement calibration constant is outside the acceptable range.
109 Over voltage cal constants out of range
The over voltage calibration constant is outside the acceptable range.
110 Wrong V+I
The instrument was unable to set the correct voltage or current value.
111 Aux vloc cal constants out of range
Calibration constants on the internal auxiliary local ADC are outside the acceptable range.
112 Aux vrem cal constants out of range
Calibration constants on the internal auxiliary remote ADC are outside the acceptable range.
113 Aux imon cal constants out of range
Calibration constants on the internal auxiliary imon ADC are outside the acceptable range.
200 Hardware error channel <channel>
A hardware error has occurred on the specified channel.
201 Invalid configuration, empty slots
There is an empty slot between modules. This configuration is not allowed.
202 Selftest Fail
A selftest failure has occurred. See selftest failure list for details.
203 Compatibility function not implemented
The requested compatibility function is not available.
204 NVRAM checksum error
A checksum error has occurred in the instrument’s nonvolatile random access memory.
205 NVRAM full
The nonvolatile random access memory of the instrument is full.
206 File not found
The internal calibration file or the internal channel attribute file was not found in NVRAM.