User`s guide
Specifications Appendix A
Model N6705A User’s Guide 123
Arbitrary Waveform Generator Maximum Bandwidth
The following tables characterize the maximum bandwidth of the
arbitrary waveform generator. The maximum bandwidth is based on
a sinewave into a resistive load and apples to any output current. The
following definitions apply in the frequency tables:
V p-p = Voltage peak-to-peak
3 dB max. = Max. frequency where the voltage drops to 3 dB below its setting
6 dB max. = Max. frequency where the voltage drops to 6 dB below its setting
THD 3 dB = The total harmonic distortion at 3 dB max. frequency
THD 3 dB = The total harmonic distortion at 6 dB max. frequency
THD < 1.5% = The frequency below which the THD is less than 1.5%.
N6751 & N6752A N6761 & N6762A
Voltage 3 dB max THD 3 dB THD < 1.5% 3 dB max THD 3 dB THD < 1.5%
0.5 Vp-p 4000 Hz 12% 440 Hz 4500 Hz 14% 450 Hz
1.0 Vp-p 2200 Hz 21% 440 Hz 3600 Hz 14% 450 Hz
2.5 Vp-p 900 Hz 25% 265 Hz 1300 Hz 25% 340 Hz
5.0 Vp-p 500 Hz 27% 160 Hz 600 Hz 25% 250 Hz
50.0 Vp-p 340 Hz 22% 25 Hz 350 Hz 22% 30 Hz
N6754A
Voltage 3 dB max THD 3 dB THD < 1.5%
0.6 Vp-p 3600 Hz 6.0% 2100 Hz
1.2 Vp-p 2600 Hz 10% 1280 Hz
3.0 Vp-p 1700 Hz 17% 800 Hz
6.0 Vp-p 1000 Hz 17% 480 Hz
60.0 Vp-p 340 Hz 22% 30 Hz
N6731B & N6741B
Voltage 3 dB max THD 3 dB 6 dB max THD 6 dB
0.1 Vp-p 175 Hz 1.0% 260 Hz 3.0%
0.1 Vp-p 125 Hz 1.0% 175 Hz 3.0%
0.3 Vp-p 75 Hz 6.0% 100 Hz 6.0%
0.5 Vp-p 40 Hz 9.0% 55 Hz 9.0%
5.0 Vp-p 20 Hz 10% 37 Hz 10%
N6732B & N6742B
Voltage 3 dB max THD 3 dB 6 dB max THD 6 dB
0.1 Vp-p 125 Hz 1.0% 200 Hz 3.0%
0.2 Vp-p 125 Hz 1.0% 180 Hz 3.0%
0.4 Vp-p 75 Hz 6.0% 100 Hz 6.0%
0.8 Vp-p 40 Hz 8.5% 60 Hz 8.5%
8.0 Vp-p 20 Hz 10% 37 Hz 10%