User`s guide
Appendix A Specifications
122 Model N6705A User’s Guide
Supplemental Characteristics (continued)
N6705A
Environmental Conditions
Operating environment
Temperature range
Relative humidity
Altitude
Storage temperature
LED statement
Indoor use, installation category II (for AC input),
pollution degree 2
0°C to 55°C (output current is derated 1% per °C above
40°C ambient temperature)
Up to 95%
Up to 2000 meters
-30°C to 70°C
Any LEDs in this unit are Class 1 LEDs as per IEC 825-1
Acoustic Noise Declaration:
This statement is provided to
comply with the requirements of
the German Sound Emission
Directive, from 18 January 1991.
Sound Pressure Lp <70 dB(A), At Operator Position,
Normal Operation, According to EN 27779 (Type Test).
Schalldruckpegel Lp <70 dB(A), Am Arbeitsplatz,
Normaler Betrieb, Nach EN 27779 (Typprüfung).
Output Terminals:
Maximum current rating
Isolation
20 A
No output terminal may be more than 240 VDC from any
other terminal or chassis ground.
BNC Trigger Connectors:
I/O
Maximum voltage
Digital TTL level compatible
5 V
USB Current Ratings:
Front panel USB connector
Rear panel USB connector
200 mA
300 mA
AC Input:
Nominal Input Ratings
Input Range
Power Consumption
Fuse
100 VAC – 240 VAC; 50/60/400Hz
86 VAC – 264 VAC
1500 VA (mainframe has power factor correction)
Internal fuse - not customer accessible.
Dimensions:
Height
Width
Depth
194.7 mm / 7.665 in.
425.6 mm / 16.756 in.
313 mm / 12.319 in.
Net Weight:
N6705A with 4 modules (typical)
Single power module (typical)
16 kg / 35 lbs
1.23 kg / 2.71 lbs