User`s guide
Specifications Appendix A
Model N6705A User’s Guide 121
Agilent N6705A DC Power Analyzer Mainframe
Supplemental Characteristics
N6705A
Maximum Output Power: (sum of total module output power)
600 W
Command Processing Time:
≤ 1 ms from receipt of command to start of output change
Protection Response Characteristics:
INH input
Fault on coupled outputs
5 μs from receipt of inhibit to start of shutdown
< 10 μs from receipt of fault to start of shutdown
Data Storage:
Internal flash memory 64 Mbytes
Digital Port Characteristics:
Maximum voltage ratings
Pins 1 and 2 as FLT output
Pins 1 - 7 as digital/trigger
outputs (pin 8 = common)
Pins 1 - 7 as digital/trigger
inputs and pin 3 as INH input
(pin 8 = common)
+16.5 VDC/− 5 VDC between pins
(pin 8 is internally connected to chassis ground).
Maximum low-level output voltage = 0.5 V @ 4 mA
Maximum low-level sink current = 4 mA
Typical high-level leakage current = 1 mA @ 16.5 VDC
Maximum low-level output voltage = 0.5 V @ 4 mA;
1 V @ 50 mA; 1.75 V @ 100 mA
Maximum low-level sink current = 100 mA
Typical high-level leakage current = 0.8 mA @ 16.5 VDC
Maximum low-level input voltage = 0.8 V
Minimum high-level input voltage = 2 V
Typical low-level current = 2 mA @ 0 V (internal 2.2k pull-up)
Typical high-level leakage current = 0.12 mA @ 16.5 VDC
Interface Capabilities:
GPIB
LXI Compliance
USB 2.0
10/100 LAN
Built-in Web server
SCPI - 1993, IEEE 488.2 compliant interface
Class C (only applies to units with LXI label on front panel)
Requires Agilent IO Library version M.01.01 or 14.0 and up
Requires Agilent IO Library version L.01.01 or 14.0 and up
Requires Internet Explorer 5+ or Netscape 6.2+
Regulatory Compliance:
EMC
Safety
Complies with EMC directive for Class A test and
measurement products.
Complies with Australian standard and carries C-Tick mark.
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 du Canada.
Electrostatic discharges greater than 1 kV near the I/O
connectors may cause the unit to reset and require
operator intervention.
Complies with European Low Voltage Directive and carries
the CE-marking.
Complies with US and Canadian safety standards for test
and measurement products.