User`s guide
Specifications Appendix A
Model N6705A User’s Guide 113
Supplemental Characteristics
N6751A / N6752A N6754A N6761A / N6762A
Programming Ranges:
Voltage high range 20 mV – 51 V 25 mV – 61.2 V 15 mV – 51 V
Voltage low range (≤ 5.5 V) N/A N/A 12 mV – 5.5 V
Current high range
10 mA – 5.1A/10 mA- 10.2A
20 mA- 20.4 A
1 mA – 1.53 A/1 mA – 3.06 A
Current low range (≤ 0.1 A) N/A N/A 0.1 mA – 0.1 A
NOTE 1
Programming Resolution:
Voltage high range 3.5 mV 4.2 mV 880 μV
Voltage low range (≤ 5.5 V) N/A N/A 90 μV
Current high range 3.25 mA 6.5 mA 60 μA
Current low range (≤ 0.1 A) N/A N/A 2 μA
Measurement Resolution:
Voltage high range 1.8 mV 2.2 mV 440 μV
Voltage low range (≤ 5.5 V) N/A N/A 44 μV
Current high range 410 μA 0.82 mA 30 μA
Current low range (≤ 0.1 A) N/A N/A 1 μA
100 μA current range (Option 1UA) N/A N/A 2 nA
200 μA current range (Option 2UA) N/A N/A 4 nA
Programming Temperature Coefficient per °C:
Voltage high range 18 ppm + 160 μV 20 ppm + 50 μV 18 ppm + 140 μV
Voltage low range (≤ 5.5 V) N/A N/A 40 ppm + 70 μV
Current high range 100 ppm + 45 μA 60ppm + 200 μA 33 ppm + 10 μA
Current low range (≤ 0.1 A) N/A N/A 60 ppm + 1.5 μA
Measurement Temperature Coefficient per °C:
Voltage high range 25 ppm + 35 μV 20 ppm + 50 μV 23 ppm + 40 μV
Voltage low range (≤ 5.5 V) N/A N/A 30 ppm + 40 μV
Current high range 60 ppm + 3 μA 60 ppm + 12 μA 40 ppm + 0.3 μA
Current low range (≤ 0.1 A) N/A N/A 50 ppm + 0.3 μA
100 μA current range (Option 1UA) N/A N/A 100 ppm + 2 nA/°C
200 μA current range (Option 2UA) N/A N/A 100 ppm + 3 nA/°C
Oscilloscope Measurement Accuracy: (@t 23 °C ±5 °C, accuracy of any individual point in the trace)
NOTE 2
Voltage 0.05% + 32 mV 0.05% + 34 mV 0.016% + 16 mV
Current high range w/Compensation on 0.1% + 14 mA 0.1% + 16 mA 0.04% + 10 mA
Current high range w/Compensation off 0.1% + 8 mA 0.1% + 16 mA 0.04% + 1 mA
Current low range w/Compensation on N/A N/A 0.03% + 10 mA
Current low range w/Compensation off N/A N/A 0.03% + 0.175 mA
Up-programming Time with full resistive load: (time from 10% to 90% of total voltage excursion)
Small voltage step 0 V to 10 V 0 V to 15 V 0 V to 10 V
Time 0.2 ms 0.35 ms 0.6 ms
Large voltage step 0 V to 50 V 0 V to 60 V 0 V to 50 V
Time 1.5 ms 2 ms 2.2 ms
1
If you are operating the unit below 255 μA in constant current mode, the output may become unregulated with the following load
conditions: The load resistance is <175 mΩ and the load inductance is >20 μH. If this occurs, an UNRegulated flag will be generated
and the output current may rise above the programmed value but will remain less than 255 μA.
2
Compensation on and Compensation off refers to the control labeled “Compensate current measurements during voltage transients”
located in the Meter View, Properties window.