User`s guide
Using the Source Functions 3
Model N6705 User’s Guide 99
From the remote interface:
The following must be observed when creating or editing an Arb
sequence:
• The Arb function type (voltage or current) must match the
Arb type specified in each sequence step.
• The Arb sequence step function waveshape must match the
shape used in the parameter commands that follow.
• The last value in the parameter list is the sequence step
number.
To set up output 1 to program a voltage sequence:
ARB:FUNC:TYPE VOLT,(@1)
ARB:FUNC:SHAP SEQ,(@1)
ARB:SEQ:RESet (@1)
To program step 0 as the voltage pulse:
ARB:SEQ:STEP:FUNC:SHAP PULS,0,(@1)
ARB:SEQ:STEP:VOLT:PULS:STAR:TIM 0.25,0,(@1)
ARB:SEQ:STEP:VOLT:PULS:TOP 10.0,0,(@1)
ARB:SEQ:STEP:VOLT:PULS:TOP:TIM 0.5,0,(@1)
ARB:SEQ:STEP:VOLT:PULS:END:TIM 0.25,0,(@1)
Note that the time of step 0 is the sum of the Start Time, Top Time
and End Time. The default pacing for sequence steps is Dwell-pacing.
Arb Trigger Sources
Specify a Trigger source for the arbitrary waveforms. The same
trigger source will be used to trigger all of the arbitrary waveforms.
From the front panel
Press the Arb key, then select the Trigger Source field.
Trigger Source:
Description:
Arb Run/Stop key
BNC Trigger in
Remote Command
The front panel Run/Stop key
The rear trigger input BNC connector
A remote interface command.