Connect the load to the output terminals using separate connecting
wires. Keep the wire-pair as short as possible and twist or bundle it
to reduce lead inductance and noise pickup. Keep the load leads
under 14.7 meters (50 feet) per lead because of inductance effects.
Connect the sense leads as close to the load as possible. Do NOT
bundle the sense wire-pair together with the load leads; keep the load
wires and sense wires separate. The sense leads carry only a few
milliamperes of current and can be a lighter gauge than the load
leads. However, any voltage drop in the sense leads can degrade the
voltage regulation of the instrument. Try to keep the sense lead
resistance less than about 0.5 per lead (this requires 20 AWG or
heavier for a 50 foot length).
Agilent Models N678xA SMU require remote sensing when using any
of the High output bandwidth modes discussed in Chapter 6. Also,
these models have additional wiring restrictions as discussed under
Agilent N678xA SMU Wiring Requirements”.
After turning the unit on, activate 4-wire remote voltage sensing by
pressing the Settings key. Navigate to and select Advanced. In the
Sense dropdown list select 4-Wire.