User`s guide

Using the Measurement Functions 4
Model N6705 User’s Guide 131
Data Logger Filename
From the front panel:
Select the Filename button to specify a filename in which to save the
data. Data will be logged to this filename the next time the Data
Logger runs. If you do not specify a filename, the data will be logged
to default.dlog, which is overwritten each time the Data Logger runs.
Enter the filename in the Path\File field. Check Append date and
time at start of log to include time-stamp information in the file.
From the remote interface:
To specify the filename in which to save the data:
INIT:DLOG “datalog1.dlog
To export a data log from the front panel display after it has been run
to a file:
MMEM:EXP:DLOG “datalog1.dlog
Data Logger Marker
From the front panel:
Select the Markers button to configure the measurements that
appear on the bottom of the display in Marker view. Measurements
apply to the portion of the trace between the two markers. You can
only select a maximum of five measurements to be displayed.