User`s guide

Using the Measurement Functions 4
Model N6705 User’s Guide 107
The Voltage Measure Only mode works best when the sense terminals
are connected to the output terminals, either by connecting the sense
and load leads to directly to the DUT, or by using the local sensing on
the front panel. Note that the input impedance in Voltage Measure
Only mode will be somewhere around 2000 pF depending on the
grounding of the DUT. This may draw up to a few microamperes of
current from the nodes that are being measured.
NOTE
The voltage measurement cannot exceed the voltage ratings of the unit as
indicated in the +/- Voltage Limit fields.
The Current Measure Only selection emulates a zero-burden
ammeter. Voltage priority mode is set by default. The +/- Current
Limits are set to their maximum values. No additional settings are
required.
NOTE
The current measurement cannot exceed the current ratings of the unit as
indicated in the +/- Current Limit fields.
When the Measure Only modes are selected, the output is enabled
and the voltage or current measurement is continuously displayed on
the Meter View as shown below. Note that both voltage and current
meter functions are active in either mode.
From the remote interface:
To specify voltage measure only:
EMUL VMET,(@1)
To measure the voltage:
MEAS:VOLT?(@1)
To specify current measure only:
EMUL AMET,(@1)
To measure the current:
MEAS:CURR?(@1)
I
measure
Current
source
+ Output
- Output
Output On
V output =
0.0000 mV