User`s guide
Table Of Contents
- Legal Notices
- Safety Notices
- In this Book
- Contents
- Quick Reference
- Installation
- Operating the Power Supply Locally
- Operating the Power Supply Remotely
- Language Reference
- SCPI Command Summary
- Calibration Commands
- Measure Commands
- Output Commands
- Source Commands
- [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
|MIN|MAX [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX][SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] |MIN|MAX [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX] - [SOURce:]CURRent:PROTection:STATe ON|OFF[SOURce:]CURRent:PROTection:STATe?
- [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
|MIN|MAX [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX][SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] |MIN|MAX [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX] - [SOURce:]VOLTage:LIMit:LOW
|MIN|MAX [SOURce:]VOLTage:LIMit:LOW? [MIN|MAX] - [SOURce:]VOLTage:PROTection:LEVel
|MIN|MAX[SOURce:]VOLTage:PROTection:LEVel? [MIN|MAX]
- [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
- Status Commands
- STATus:PRESet
- STATus:OPERation[:EVENt]?
- STATus:OPERation:CONDition?
- STATus:OPERation:ENABle
STATus:OPERation:ENABle? - STATus:OPERation:NTR
STATus:OPERation:PTR STATus:OPERation:NTR? STATus:OPERation:PTR? - STATus:QUEStionable[:EVENt]?
- STATus:QUEStionable:CONDition?
- STATus:QUEStionable:ENABle
STATus:QUEStionable:ENABle? - STATus:QUEStionable:NTR
STATus:QUEStionable:PTR STATus:QUEStionable:NTR? STATus:QUEStionable:PTR? - *CLS
- *ESE*ESE?
- *ESR?
- *OPC*OPC?
- *SRE*SRE?
- *STB?
- *WAI
- System Commands
- Trigger Commands
- Programming Examples
- Specifications
- Verification and Calibration
- Verification
- Equipment Required
- Measurement Techniques
- Constant Voltage Tests
- Constant Current Tests
- Test Record Form – Agilent N5741A and N5761A
- Test Record Form – Agilent N5742A and N5762A
- Test Record Form – Agilent N5743A and N5763A
- Test Record Form – Agilent N5744A and N5764A
- Test Record Form – Agilent N5745A and N5765A
- Test Record Form – Agilent N5746A and N5766A
- Test Record Form – Agilent N5747A and N5767A
- Test Record Form – Agilent N5748A and N5768A
- Test Record Form – Agilent N5749A and N5769A
- Test Record Form – Agilent N5750A and N5770A
- Test Record Form – Agilent N5751A and N5771A
- Test Record Form – Agilent N5752A and N5772A
- Calibration
- Verification
- Service
- Compatibility
- Index
- Declaration of Conformity

Appendix A Specifications
88 Series N5700 User’s Guide
Performance Specifications
Agilent Models N5741A – N5752A and N5761A – N5772A
Model
N5741A
N5761A
N5742A
N5762A
N5743A
N5763A
N5744A
N5764A
N5745A
N5765A
N5746A
N5766A
N5747A
N5767A
N5748A
N5768A
N5749A
N5769A
N5750A
N5770A
N5751A
N5771A
N57521A
N57721A
DC Output Ratings:
Voltage 6V 8V 12.5V 20V 30V 40V 60V 80V 100V 150V 300V 600V
Current 750W 100A 90A 60A 38A 25A 19A 12.5A 9.5A 7.5A 5A 2.5A 1.3A
Current 1500W 180A 165A 120A 76A 50A 38A 25A 19A 15A 10A 5A 2.6A
Power 750W 600W 720W 750W 760W 750W 760W 750W 760W 750W 750W 750W 780W
Power 1500W 1080W 1320W 1500W 1520W 1500W 1520W 1500W 1520W 1500W 1500W 1500W 1560W
Output Ripple and Noise:
CV p-p
NOTE 1
60mV 60mV 60mV 60mV 60mV 60mV 60mV 80mV 80mV 100mV 150mV 300mV
CV rms
NOTE 2
8mV 8mV 8mV 8mV 8mV 8mV 8mV 8mV 8mV 12mV 20mV 60mV
Load Effect: (change from 10% to 90% of full load)
Voltage 2.6mV 2.8mV 3.25mV 4mV 5mV 6mV 8mV 10mV 12mV 17mV 32mV 62mV
Current 750W 25mA 23mA 17mA 12.6mA 10mA 8.8mA 7.5mA 6.9mA 6.5mA 6mA 5.5mA 5.26mA
Current 1500W 41mA 38mA 29mA 20.2mA 15mA 12.6mA 10mA 8.8mA 8mA 7mA 6mA 5.5mA
Source Effect: (change from 85-132 VAC input or 170-265 VAC input)
Voltage 2.6mV 2.8mV 3.25mV 4mV 5mV 6mV 8mV 10mV 12mV 17mV 32mV 62mV
Current 750W 12mA 11mA 8mA 5.8mA 4.5mA 3.9mA 3.25mA 2.95mA 2.75mA 2.5mA 2.25mA 2.13mA
Current 1500W 20mA 18.5mA 14mA 9.6mA 7mA 5.8mA 4.5mA 3.9mA 3.5mA 3mA 2.5mA 2.26mA
Programming Accuracy:
Voltage 0.05%+ 3mV 4mV 6.25mV 10mV 15mV 20mV 30mV 40mV 50mV 75mV 150mV 300mV
Current 750W 0.1%+ 100mA 90mA 60mA 38mA 25mA 19mA 12.5mA 9.5mA 7.5mA 5mA 2.5mA 1.3mA
Current 1500W 0.1%+ 180mA 165mA 120mA 76mA 50mA 38mA 25mA 19mA 15mA 10mA 5mA 2.6mA
Measurement Accuracy:
Voltage 0.1%+ 6mV 8mV 12.5mV 20mV 30mV 40mV 60mV 80mV 100mV 150mV 300mV 600mV
Current 750W 0.1%+ 300mA 270mA 180mA 114mA 75mA 57mA 37.5mA 28.5mA 22.5mA 15mA 7.5mA 3.9mA
Current 1500W 0.1%+ 540mA 495mA 360mA 228mA 150mA 114mA 75mA 57mA 45mA 30mA 15mA 7.8mA
Load Transient Recovery Time:
(time for output voltage to recover within 0.5% of its rated output for a load change from 10 to 90% of its rated output current)
Time ≤ 1.5ms ≤ 1.5ms ≤ 1.5ms ≤ 1ms ≤ 1ms ≤ 1ms ≤ 1ms ≤ 1ms ≤ 1ms ≤ 2ms ≤ 2ms ≤ 2ms
Voltage set point from 10% to 100% of rated output
NOTE 1
20MHz
NOTE 2
From 5Hz - 1MHz