User`s guide
Table Of Contents
- Legal Notices
- Safety Notices
- In this Book
- Contents
- Quick Reference
- Installation
- Operating the Power Supply Locally
- Operating the Power Supply Remotely
- Language Reference
- SCPI Command Summary
- Calibration Commands
- Measure Commands
- Output Commands
- Source Commands
- [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
|MIN|MAX [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX][SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] |MIN|MAX [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX] - [SOURce:]CURRent:PROTection:STATe ON|OFF[SOURce:]CURRent:PROTection:STATe?
- [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
|MIN|MAX [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX][SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] |MIN|MAX [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX] - [SOURce:]VOLTage:LIMit:LOW
|MIN|MAX [SOURce:]VOLTage:LIMit:LOW? [MIN|MAX] - [SOURce:]VOLTage:PROTection:LEVel
|MIN|MAX[SOURce:]VOLTage:PROTection:LEVel? [MIN|MAX]
- [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
- Status Commands
- STATus:PRESet
- STATus:OPERation[:EVENt]?
- STATus:OPERation:CONDition?
- STATus:OPERation:ENABle
STATus:OPERation:ENABle? - STATus:OPERation:NTR
STATus:OPERation:PTR STATus:OPERation:NTR? STATus:OPERation:PTR? - STATus:QUEStionable[:EVENt]?
- STATus:QUEStionable:CONDition?
- STATus:QUEStionable:ENABle
STATus:QUEStionable:ENABle? - STATus:QUEStionable:NTR
STATus:QUEStionable:PTR STATus:QUEStionable:NTR? STATus:QUEStionable:PTR? - *CLS
- *ESE*ESE?
- *ESR?
- *OPC*OPC?
- *SRE*SRE?
- *STB?
- *WAI
- System Commands
- Trigger Commands
- Programming Examples
- Specifications
- Verification and Calibration
- Verification
- Equipment Required
- Measurement Techniques
- Constant Voltage Tests
- Constant Current Tests
- Test Record Form – Agilent N5741A and N5761A
- Test Record Form – Agilent N5742A and N5762A
- Test Record Form – Agilent N5743A and N5763A
- Test Record Form – Agilent N5744A and N5764A
- Test Record Form – Agilent N5745A and N5765A
- Test Record Form – Agilent N5746A and N5766A
- Test Record Form – Agilent N5747A and N5767A
- Test Record Form – Agilent N5748A and N5768A
- Test Record Form – Agilent N5749A and N5769A
- Test Record Form – Agilent N5750A and N5770A
- Test Record Form – Agilent N5751A and N5771A
- Test Record Form – Agilent N5752A and N5772A
- Calibration
- Verification
- Service
- Compatibility
- Index
- Declaration of Conformity
Operating the Power Supply Locally 3
Series N5700 User’s Guide 35
UVL Check
• Press the OVP/UVL button twice so that the DC AMPS display
indicates UUL. The DC VOLTS display shows the UVL level.
• Use the voltage knob and set the UVL level of the unit to 50%
of its full-scale voltage rating or 30 volts, whichever is lower.
• Wait a few seconds until the DC VOLTS display returns to
show the output voltage.
• Use the voltage knob and lower the output voltage of the unit
until it approaches the UVL setting. Check to make sure that
the output voltage cannot be set lower than the UVL setting.
• Press the OVP/UVL button twice. Rotate the voltage knob and
reset the UVL level of the unit to its minimum setting.
Constant Current Check
• Turn the POWER switch off. Wait a few seconds until the AC
indicator on the front panel goes out.
• Use a heavy wire and short the +V and –V output terminals
together.
• Turn the POWER switch on.
• Turn the output on by pressing the OUT ON button. The
green OUT ON indicator should be illuminated. The green CC
indicator should be also illuminated.
• Rotate the current knob while observing the DC AMPS
display. The output current should vary while the knob is
turned. The current range is from zero to the maximum rated
output for the power supply model.
OCP Check
• Rotate the current knob and set the current limit of the unit
to about 10% of its full-scale current rating.
• Press the OCP button. This should trip the OCP protection.
The OCP indicator should be illuminated, the DC VOLTS
display should indicate OCP, and the Alarm indicator should
be blinking.
• Press the OCP button again to cancel OCP protection. The DC
VOLTS display should indicate OFF because the OCP
protection is latched.
• Press the OUT ON button to reset the OCP protection. The
output should return to its previous setting.
• Turn the POWER switch off.
• Remove the short from the +V and –V output terminals.