Technical data
62 Notes on USB Electrical Compliance Testing
3 Device Hi-Speed Tests
EL_8 K Test
The USB automated test application will prompt you to perform these
steps:
1 Return the Test switch of the test fixture to the NORMAL position.
2 Cycle the device power to restore the device to normal operation. On
the Device Test Menu of the HS Electrical Test Tool, click [Enumerate
Bus] once.
3 Select TEST_K from the Device Command drop down menu. Click
[EXECUTE] once to place the device into TEST_K test mode.
4 Switch the test fixture into the TEST position.
5 Click OK to close the Test Instructions dialog.