Technical data

Agilent Technologies
141
N5416A USB 2.0 Compliance Test Option
Notes on Electrical Testing
5
Host Hi-Speed Electrical Tests
Host Hi-Speed Signal Quality 143
Host Controller Packet Parameters 148
Host Disconnect Detect 155
Host CHIRP Timing 159
Host Suspend/Resume Timing 164
Host Test J/K, SE0_NAK 169
In addition to the hi- speed electrical tests prescribed in this chapter, the
host controller under test must also pass the following electrical
compliance tests applicable to the EHCI Host Controller:
Full speed signal quality.
Low speed signal quality.
Drop/Droop.
Before Running These Tests
If you haven’t already performed the initial equipment set up, see “Setting
Up the Equipment" on page 25.
NOTE
To give the automated test software more flexibility in making corner case measurements
(and improve test performance), the oscilloscope’s memory upgrade option is
recommended (Option 001 when ordered with the oscilloscope, or after purchase: E2680A
for the 5485xA Series oscilloscopes, or N5404A for the 80000 and 90000A Series
oscilloscopes).
Also, to reduce the bandwidth of the oscilloscope to just what is required by the USB-IF
(2 GHz for USB 2.0)—this can greatly reduce system measurement noise, allowing the
maximum measurement margin—the enhanced bandwidth software option is
recommended for the Agilent 54855A oscilloscope (Option 008), and the noise reduction
software option is recommended for the 80000 and 90000A Series oscilloscopes (Option
005 when ordered with the oscilloscope or N5403A after purchase).