U3025AE10 Operational Check
4. In the “Electronic Calibration Step 1 of 1” window, press Measure.
5. At the Calibration Completed prompt, select Save As User Calset and type the name
999.1. Overwrite the Calset if it already exists. Press Save.
NOTE If you do not have a key board, select Save As User Calset > Edit Name and
save as 999.x. X is the port number you are calibrating. See Figure 45.
Figure 45 Calibration Complete
6. Repeat step 1 thru step 5 (1-Port Calibration Procedure) for Ports 2 thru 12. When
finished, there should be twelve Cal Sets saved with the titles “999.1” thru “999.12”
(12-Port).
If you are using an ECal module you can verify the individual port calibration by
selecting Calibration > ECal Confidence Check. For further information refer to the
system Help menu.
7. Press Trace > Delete Trace. There should be no traces on the PNA screen.
8. Select Calibration > Cal Set Viewer to launch the Cal Set Viewer toolbar. Refer to
Figure 46 on page 49.