24 User’s Guide
U3025AE10 Controlling the System with E8363/4B and N5230A
Trace Measure S-Parameter
S-Parameter selection can be accomplished using Trace Measure menu. Select TRACE >
Measure. Use the drop-down menus to select 1 of 144 S-Parameters for the 12-Port system,
see Figure 17. The first number in the Sxx selection is the Receiver Port and the second
number will be the Source Port. Any port can be selected to be the Receiver, Source or both,
as in S11. The front panel R LED indicates the port is the Receiver and the S LED
indicates the port is the Source.
Figure 17 12-Port Trace Measure
New Trace Measure S-Parameter
S-Parameter Tab: Multiple S-Parameters can be made from the New Measurement
menu. In the drop-down menu select Trace > New Trace. The New Measurement window
allows the selection of any of the 144 S-Parameter's. SeeFigure 18 and Figure 19.
Figure 18 12-Port New Trace Measure (S11 - S88)