User`s guide
User’s Guide 23
U3025AE10 Controlling the System with E8363/4B and N5230A
External Test Set Control Feature
To verify that the network application has the U302xAE10 interface features, select
Channel > External Test Set > Other. The U302xAE10 will be displayed as External Test
Set Control-U302xAE10. See Figure 15 or Figure 16. The U3022AE10 test set file can be
used for the U3025AE10 file.
This menu will allow the physical Ports 1 through 12 to be identified as any port for your
convenience. For example; Port 5 can be named Port 2.
The External Test Set Control-U302xAE10 also allows control of the DUT control lines,
refer to “DUT Control Lines” on page 38. To change the state from LOW to HIGH, select
the graphical user interface (GUI) for the specific control (LINE 1 through 8) and then
press OK. Each line can be controlled separately, see Figure 15.
Figure 15 External Test Set U302xAE10 (Port 1 - 8)
Select the Port Control down arrow for Ports 9 through 12, see Figure 16.
Figure 16 External Test Set U302xAE10 (Port 9 - 12)
Control Lines