User`s guide
User’s Guide 7
U3025AE10 General Specifications
General Specifications
Specifications for the U3025AE10 Multiport Test Set (10 MHz to 50 GHz) are
characteristic for the System performance of the PNA and Test Set. Actual performance of
the system is based on the customers PNA that is used with the test set. A functional
certificate is only offered for the U3025AE10.
NOTE When connected to a PNA, this test set will degrade the performance at the
test ports. The internal switch paths reduce test port power and power to the
receivers. This affects the test port power of the PNA and also reduces
dynamic range. The reflection tracking values measured in the “Operational
Check” on page 43 can be subtracted from the analyzers dynamic range to
determine the approximate performance of the system.
Power Requirements
Verify that the required ac power is available at all necessary locations before installing
the Test Set to the PNA.
• Air conditioning equipment (or other motor–operated equipment) should not be placed
on the same ac line that powers the Test Set and PNA.
• Table 2 contains the maximum wattage for all instruments. This table can be use to
determine the electrical and cooling requirements.
NOTE Values are based on 120 Vac supplied to each instrument at 60 Hz.
WARNING This is a Safety Class I product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall be
inserted only into a socket outlet provided with a protective earth
contact. Any interruption of the protective conductor, inside or
outside the instrument, is likely to make the instrument dangerous.
Intentional interruption is prohibited.
Table 2 Power Requirements
Standard Equipment
Instrument Maximum
Wattage
E8364B/C and
N5230A/C
350
U3025AE10 350
Total 700