Agilent Technologies U3025AE10 User’s Guide Use this manual with the following documents: PNA Series Network Analyzer On-line Help System and Application Note 1408-12 Manufacturing Part Number: U3025-90001 Printed Date: April 2008 Supersede: July 2007 © Copyright 2007, 2008 Agilent Technologies, Inc. All rights reserved.
Warranty Statement THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Safety Notes The following safety notes are used throughout this document. Familiarize yourself with each of these notes and its meaning before performing any of the procedures in this document. WARNING Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met. CAUTION Caution denotes a hazard.
iv User’s Guide
Contents U3025AE10 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Network Analyzer Requirement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Available Options . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents DUT Control Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 Setting the DUT Control Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 Setting the Variable Source Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40 Connecting to the DUT Control Lines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
U3025AE10 User’s Guide 1
U3025AE10 Introduction Introduction This document describes how to use the U3025AE10 Multiport Test Set with the Agilent E8363/4B/C 2-Port and the N5230A/C 2-Port PNA Network Analyzer.
U3025AE10 Description Description The Agilent U3025AE10 is a 10-Port mechanical switching extension test set (10 MHz to 50 GHz). The 10-Port extension test set can be connected to a 12-Port device when configured with the E8364B/C Option 014 or N5230A/C Option 525 or 425 2-Port network analyzers. The U3025AE10 can be used with the E8363B/C (10 MHz to 40 GHz). The test ports are 2.4 mm male. The test set is controlled by the test set I/O connector located on the rear panel of the PNA network analyzer.
U3025AE10 Network Analyzer Requirement Network Analyzer Requirement • All PNA, PNA-L Network Analyzers require Option 551, which adds N-Port error correction and full cross-bar measurement capability. • PNA firmware revision: N5230A and E8363/4B Option 551 ≥ A.07.50.26 N5230C and E8363/4C Option 551 ≥ A.08.04.10 • The N5230A/C PNA-L requires Option 525 or 425 (40 GHz) to provide the test set interface connections and requires test set file u302xae10_p2.tsx.
U3025AE10 Available Options Available Options Test Set Options The U3025AE10 has two available test set options: Refer to “System Block Diagrams” beginning on page 58. • Standard - Includes amplifiers. • Option 001 provides Bias Tee’s for the 10-Ports in the U3025AE10. Accessory Options Installation instructions are included in the option package. • Option 1CM - Rackmount Kit (5063-9215). • Option 1CN - Front-Handle Kit (5063-9228). • Option 1CP - Rackmount with front handle Kit (5063-9222).
U3025AE10 Verifying the Shipment Verifying the Shipment Inspect the shipping container. If the container or packing material is damaged, it should be kept until the contents of the shipment have been checked mechanically and electrically. If there is physical damage refer to “Contacting Agilent” on page 65. Keep the damaged shipping materials (if any) for inspection by the carrier and an Agilent Technologies representative. Table 1 contains the accessories shipped with your U3025AE10.
U3025AE10 General Specifications General Specifications Specifications for the U3025AE10 Multiport Test Set (10 MHz to 50 GHz) are characteristic for the System performance of the PNA and Test Set. Actual performance of the system is based on the customers PNA that is used with the test set. A functional certificate is only offered for the U3025AE10. When connected to a PNA, this test set will degrade the performance at the test ports.
U3025AE10 General Specifications Environmental Requirements The environmental requirements of the PNA are listed in Table 3. Note that these requirements are the same as those of the E8364B/C and N5230A/C. Ventilation Requirements: When installing the instrument in a cabinet, the convection into and out of the instrument must not be restricted.
U3025AE10 General Specifications Dimensions and Space Requirements Standard installation of the U3025AE10 and PNA includes configuration and installation on a customer provided lab bench or table top of adequate size and strength.
U3025AE10 General Specifications Frequency Range and Maximum Power Levels The U3025AE10 frequency range is 10 MHz to 50 GHz. CAUTION Table 6 It is recommend that you do not operate components near damage levels (+30 dBm). The power levels must be 3 dB below maximum level to ensure no damage. See Table 6.
U3025AE10 Table 7 General Specifications Typical Reflection Tracking Loss1 2 Standard (dB) (typical) Option 001 (dB) (typical) 10 MHz to 45 MHz +3.0 2.5 45 MHz to 500 MHz +6.5 +6.0 500 MHz to 2 GHz +4.0 3.5 2 GHz to 10 GHz +2.0 1.5 10 GHz to 20 GHz 0 −0.5 20 GHz to 30 GHz −3.0 −8.0 30 GHz to 40 GHz −4.5 −9.0 40 GHz to 45 GHz −7.0 −10.0 45 GHz to 50 GHz −8.5 −12.0 Frequency 1. Reflection tracking loss is generally source path loss + receiver path loss. 2.
U3025AE10 General Specifications Front and Rear Panel Features Figure 3 Front Panel (Multiport Test Set) Access Ports Access Ports Active LED Test Ports Test Port LEDs Active LED • When the test set power switch is On and is connected and addressed by a PNA, the LED is On (illuminated). • The LED is Off (not illuminated) when the test set power switch is in Standby, or not addressed by a PNA. Test Port LEDs • An illuminated LED indicates an active port and if it is in Source or Receiver mode.
U3025AE10 General Specifications Access Ports – 2.4 mm (female) • SOURCE OUT • CPLR THRU • CPLR ARM • A IN – B IN Line Switch – Standby 1 – ON (Active LED On) Figure 4 Rear Panel (Multiport Test Set) Control Lines Voltage Adjust Option 001 Bias Tee Inputs Line Module Test Set Interface Pass Through Interface Control Lines and Voltage Adjust For further information pertaining to control lines and voltage adjustments see “DUT Control Lines” on page 38.
U3025AE10 General Specifications Line Module The line fuse, as well as a spare, reside within the line module. Figure 5 illustrates where the fuses are located and how to access them. Available Fuses • Fuse (F 5 A/250V, 2110-0709) U.L. listed and CSA certified WARNING For continued protection against fire hazard replace line fuse only with same type and rating: • F 5A/250V, Part Number 2110-0709 The use of other fuses or material is prohibited.
U3025AE10 System Setup E8363B/C, E8364B/C, or N5230A/C System Setup E8363B/C, E8364B/C, or N5230A/C The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened. WARNING Preparing the PNA Network Analyzer 1. Remove the feet from the bottom of the network analyzer. Refer to Figure 6. 2. Remove the 2 lower standoffs and screws (0515-1619) from the rear panel on the network analyzer. Refer to Figure 6.
U3025AE10 System Setup E8363B/C, E8364B/C, or N5230A/C 4. Place the network analyzer on top of the Test Set and ensure that the front frame of the network analyzer is positioned slightly forward of the locks that are attached to the Test Set. Slide the network analyzer back so the locks engage the front frame of the analyzer. Refer to Figure 8.
U3025AE10 System Setup E8363B/C, E8364B/C, or N5230A/C 12-Port RF Cable Connections Figure 10 on page 18 illustrates the setup configuration of the U3025AE10 Multiport Test Set and how it should be configured to the E8364B/C 2-Port and N5230A/C PNA Network Analyzer. 1. Connect the RF cables supplied with this option between the network analyzer and the U3025AE10 Test Set. Torque each cable to 8 in-lb. Refer to Table 8 and Figure 10 on page 18.
U3025AE10 Figure 10 System Setup E8363B/C, E8364B/C, or N5230A/C 12-Port Setup Configuration 2. Connect the PNA Test Set I/O cable (8120-6818) to the U3025AE10 Test Set Interface on the rear panel.
U3025AE10 Controlling the System with E8363/4B and N5230A Controlling the System with E8363/4B and N5230A This section will describe how to setup and operate the U3025AE10 Multiport Test Set with the E8364B 2-Port PNA Network Analyzer or N5230A PNA-L. The internal firmware of the Agilent E8363/4B/C 2-Port Network Analyzer has not been modified for this test set option. Power levels may differ from those indicated on the PNA when the Test Set is connected.
U3025AE10 Controlling the System with E8363/4B and N5230A Installing New Firmware Before installing new firmware into your PNA-L, you must store the mixer calibration files. 1. From the Start menu select run and enter 1 > OK. 2. Command line “Enter a Choice:”, using the front panel keys enter [6]. 3. Command line “Press any key to continue....”, use any front panel key.
U3025AE10 Controlling the System with E8363/4B and N5230A 4. Copy the following three mxcalfiles from D:\CALFILES to c:\Program Files\Agilent\Network Analyzer.
U3025AE10 Controlling the System with E8363/4B and N5230A 12 Port System with E8363/4B or N5230A The Option 551 must be installed for Multiport capability. To access the multiport application select System > Configure > Multiport Capability. See Figure 13. Figure 13 System Configuration Select U302xAE10 (12-Port System) from the test set drop-down menu and select Restart as a Multiport PNA with this test set. Press OK.
U3025AE10 Controlling the System with E8363/4B and N5230A External Test Set Control Feature To verify that the network application has the U302xAE10 interface features, select Channel > External Test Set > Other. The U302xAE10 will be displayed as External Test Set Control-U302xAE10. See Figure 15 or Figure 16. The U3022AE10 test set file can be used for the U3025AE10 file. This menu will allow the physical Ports 1 through 12 to be identified as any port for your convenience.
U3025AE10 Controlling the System with E8363/4B and N5230A Trace Measure S-Parameter S-Parameter selection can be accomplished using Trace Measure menu. Select TRACE > Measure. Use the drop-down menus to select 1 of 144 S-Parameters for the 12-Port system, see Figure 17. The first number in the Sxx selection is the Receiver Port and the second number will be the Source Port. Any port can be selected to be the Receiver, Source or both, as in S11.
U3025AE10 Figure 19 Controlling the System with E8363/4B and N5230A 12-Port New Trace Measure (S88 - S12_12) Scroll Bar Balanced Tab: Balanced Measurements can be configured by selecting the Balance tab in the New Measurement menu. Refer to Figure 20. For more information on balanced (differential) component measurement, refer to the Application Note 1373-1 and 1373-2 (5988-5634EN and 5988-5635EN) at http://www.home.agilent.com. In the search menu type “Multiport and Balanced.
U3025AE10 Controlling the System with E8363/4B and N5230A Receiver Tab: The S-Parameter measurements can be ratioed with selectable Denominators for each port and receiver. Refer to the standard PNA documentation for more information.
U3025AE10 Controlling the System with E8363/4B and N5230A N-Port Calibration with E8363/4B and N5230A It is recommended that you perform an ECal characterization to minimize the connections required for multiple port calibration. The N4693A Option M0F is recommended with cable (85133F) if you are calibrating at the PNA and test set ports. Characterize the ECal module with adapters that will not be used in the measurement of the DUT.
U3025AE10 Controlling the System with E8363/4B and N5230A 2. Continue following the Cal Wizard prompts. On the “Select Calibration Ports and ECal Module” window, press Select All or select the ports you are calibrating and press Next. Figure 23 12-Port Calibration Help 3. Connect the ECal or the mechanical cal kit to the ports you are calibrating following the Cal Wizard prompts and press Measure after each connection. 4.
U3025AE10 Controlling the System with E8363/4C and N5230C Controlling the System with E8363/4C and N5230C This section will describe how to setup and operate the U3025AE10 Multiport Test Set with the E8363/4C and N5230C 2-Port PNA Network Analyzer. The U3025AE10 Multiport Test Set is considered a “slave” instrument. A E8363/4C or N5230C must be used to control the Test Set. • PNA Multiport Mode Option 551 requires firmware revision ≥ A.08.04.10.
U3025AE10 Controlling the System with E8363/4C and N5230C PNA Multiport Mode with Option 551 The PNA Multiport Mode selects the test set file that will enable the PNA to control the test set. The PNA Multiport Mode allows you to complete a N-Port calibration using the Cal Wizard application in the PNA. Refer to the PNA Help system for more information. 12-Port System with E8363/4C and N5230C The E8363/4C and N5230C Option 551 must be installed for Multiport capability.
U3025AE10 Controlling the System with E8363/4C and N5230C External Test Set Control Feature To verify that the network application has the U3025AE10 interface features, select Channel > External Test Set > Other. The U3025AE10 will be displayed as External Test Set Control-U3025AE10 (U302xAE10 or U3022AE10). Figure 27 External Test Set Control This menu will allow the physical Ports 1 through 12 to be identified as any port for your convenience. For example; Port 5 can be named Port 2.
U3025AE10 Controlling the System with E8363/4C and N5230C The External Test Set Control-U3025AE10 also allows control of the DUT control lines, refer to “DUT Control Lines” on page 38. To change the state from LOW to HIGH, select the graphical user interface (GUI) for the specific control (LINE 1 through 8) and then press OK. Each line can be controlled separately, see Figure 28. Figure 28 External Test Set U3025AE10 (Port 1 - 8) Control Lines Select the Port Control down arrow for Ports 9 through 12.
U3025AE10 Controlling the System with E8363/4C and N5230C Trace Measure S-Parameter S-Parameter selection can be accomplished using Response Measure menu. Select TRACE > Measure. Use the drop-down menus to select 1 of 144 S-Parameters for the 12-Port system, see Figure 30. The first number in the Sxx selection is the Receiver Port and the second number will be the Source Port. Any port can be selected to be the Receiver, Source or both, as in S11.
U3025AE10 Controlling the System with E8363/4C and N5230C Figure 32 12-Port New Trace Measure (S11 - S88) Figure 33 12-Port New Trace Measure (S88 - S12_12) Scroll Bar 34 User’s Guide
U3025AE10 Controlling the System with E8363/4C and N5230C Balanced Tab: Balanced Measurements can be configured by selecting the Balance tab in the New Measurement menu. For more information on balanced (differential) component measurement, refer to the Application Note 1373-1 and 1373-2 (5988-5634EN and 5988-5635EN) at http://www.home.agilent.com. In the search menu type in “Multiport and Balanced.
U3025AE10 Controlling the System with E8363/4C and N5230C N-Port Calibration with E8363/4C and N5230C It is recommended that you perform an ECal characterization to minimize the connections required for multiple port calibration. The N4693A Option M0F is recommended with cable (85133F) if you are calibrating at the PNA and test set ports. Characterize the ECal module with adapters that will not be used in the measurement of the DUT.
U3025AE10 Controlling the System with E8363/4C and N5230C 2. Continue following the Cal Wizard prompts. On the “Select Calibration Ports and ECal Module” window, press Select All or select the ports you are calibrating and press Next. Figure 37 12-Port Calibration Help 3. Connect the ECal or the mechanical cal kit to the ports you are calibrating following the Cal Wizard prompts and press Measure after each connection. 4.
U3025AE10 DUT Control Lines DUT Control Lines The 15 pin female D-Sub connector on the rear panel provides 8 latched data connections that can be used to control your DUT. An adjustable voltage source (+2 to +5 Vdc) is provided on the front panel. A positive or negative external source can be used. Refer to Table 11 and Figure 41 on page 42. Setting the DUT Control Interface This section describes how to control the DUT control lines.
U3025AE10 DUT Control Lines NOTE All DUT control lines must be set with each command sent.
U3025AE10 DUT Control Lines Setting the Variable Source Voltage The output voltage of pin 12 can be varied from +2 to +5 V. Perform the following procedure to set the voltage: 1. Turn On U3025AE10. 2. Measure the voltage between pin 12 and 15 using a multimeter. 3. Rotate the voltage adjustment trimmer on the front panel until the multimeter indicates the appropriate voltage.
U3025AE10 DUT Control Lines Using an External Power Supply Figure 39 illustrates an example of the connection between the DUT and the U3025AE10 with an external dc power supply. Input the High and Low signals from the external power supply to the Positive Input and Negative Input respectively, and connect each line to the control terminal of the DUT. Turning On the U3025AE10 using an External Power Supply. 1. Turning On the U3025AE10. 2. Connect the DUT. 3. Turn On the external power supply.
U3025AE10 Table 11 DUT Control Lines DUT Control Specifications Item Specifications Connector Shape 15–pin female D–Sub Voltage Range: Figure 41 42 Positive Input 0 to +5 V Negative Input –5 to 0 V Maximum Current 100 mA (in total of each line) Impedance < 10 Ω Range of Variable Voltage +2 to +5 V Block Diagram of DUT Control User’s Guide
U3025AE10 Operational Check Operational Check This section provides the test calibration procedure to confirm the U3025AE10 and PNA operational performance. The operation verification limits provided ensure that your U3025AE10 and PNA are operating properly. Verification Limits Specifications for the U3025AE10 Multiport Test Set are typical. System performance for the PNA and Test Set are only characteristic and intended as non warranted information.
U3025AE10 Operational Check Equipment Required The Agilent U3025AE10 requires that the user be familiar with the equipment and components listed in Table 13. This section provides an equipment list and setup of the PNA and Test Set. Table 13 Equipment List Description Qty E8364B/C 2-Port Network Analyzer (Option 014 or 551) or N5230A/C 2-Port Network Analyzer (Option 525 or 551) 1 N4693A ECal Module with a female connector, or a 2.4 mm Cal kit (85056B, 85057D, etc.
U3025AE10 Operational Check Operational Check Procedure with E8363/4B or N5230A The sequence of this procedure is very important and must be followed or the performance accuracy and results may vary from the reference plots provided. Prepare the PNA 1. Connect the Test Set to the E836/34B or N5230A using the interconnect cables as shown in Figure 10 on page 18 and Table 8 on page 17. 2. On the PNA, press Calibration > Cal Sets. Delete or Rename any Cal Sets titled “999.1” thru “999.
U3025AE10 Figure 42 Operational Check Test Set to PNA Connections 11.Allow the ECal module, Test Set and PNA to warm up for a minimum of 30 minutes.
U3025AE10 Operational Check 1-Port Calibration Procedure 1. Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing. Torque to 8 in-lb. For more information press the Help button, see Figure 43. Figure 43 ECal at Port 1 2. Perform a 1-Port Calibration on Port 1. On the PNA, press Calibration > Calibration Wizard. a. If using a mechanical cal kit, select SmartCal. b. If using an ECal module, select ECal. 3. Continue following the Cal Wizard prompts.
U3025AE10 Operational Check 4. In the “Electronic Calibration Step 1 of 1” window, press Measure. 5. At the Calibration Completed prompt, select Save As User Calset and type the name 999.1. Overwrite the Calset if it already exists. Press Save. NOTE Figure 45 If you do not have a key board, select Save As User Calset > Edit Name and save as 999.x. X is the port number you are calibrating. See Figure 45. Calibration Complete 6.
U3025AE10 Figure 46 Operational Check Calibration, Cal Set Viewer 9. From the Cal Sets drop-down menu, select 999.x (the next sequential number). Select the Reflection Tracking(1,1) term in the center drop-down menu and ensure that the Enable and Error Terms are selected as shown in Figure 47. NOTE Figure 47 You may also create a table on the PNA and enter the limit line stimulus and response values. Select Limit Test ON and Limit Line ON, and press Show Table. See Figure 47.
U3025AE10 Operational Check 10.Compare the Reflection Tracking (1,1) trace to the appropriate limits in Table 12 on page 43. This can be done using Limit Lines (press Trace > Limit Test) or Marker. The trace should be above the limit line and PASS will be displayed on the screen. Refer to Figure 48. 11.Repeat step 9 and step 10 for Cal Sets “999.1” thru “999.12” (12-Port).
U3025AE10 Operational Check Operational Check Procedure with E8363/4C and N5230C 1. Connect the Test Set to the 2-Port E8363/4C or N5230C 2-Port PNA as shown in Figure 10, Table 11 on page 42, Table 8 and Table 8 on page 17. 2. Turn On the Test Set. 3. On the E8363/4C and N5230C, press Response > CAL > Manage CALS > CAL Set. Delete or Rename any Cal Sets titled “999.1” thru “999.12” (12-Port), although it is unlikely that you will find Cal Sets with these names. 4. Verify that the PNA is in 12-Port.
U3025AE10 Operational Check Operational Check Procedure for E8363/4C and N5230C Systems 1. Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing. Torque to 8 in-lb. For more information press the Help button, see Figure 50. 2. Perform a 1-Port Calibration on Port 1. On the PNA, press Response > CAL > Start Cal > Calibration Wizard. a. If using a mechanical cal kit, select SmartCal. b. If using an ECal module, select ECal. 3. Continue following the Cal Wizard prompts.
U3025AE10 Operational Check 5. At the Calibration Completed prompt, select Save As User Calset and type the name 999.1. Overwrite the Calset if it already exists. Press Save. NOTE If you do not have a key board, select Save As User Calset > Edit Name and save as 999.x. X is the port number you are calibrating. See Figure 51. Use the numeric keypad on the PNA front panel to enter "999.1." Figure 51 Calibration Complete 6. Repeat step 1 thru step 5 (1-Port Calibration Procedure) for Ports 2 thru 12.
U3025AE10 Operational Check 8. To launch the Cal Set Viewer toolbar. Select Response > CAL > Manage CALS > CAL Set Viewer. Figure 52 Calibration, Cal Set Viewer 9. From the Cal Sets drop-down menu, select 999.1 and check Enable. Select the Reflection Tracking(1,1) term in the center drop-down menu and ensure that the Enable and Error Terms are selected. NOTE You may also create a table on the network analyzer and enter the limit line stimulus and response values.
U3025AE10 Figure 54 Operational Check Reflection Tracking Trace (Port 1-2) 10.Compare the Reflection Tracking (1,1) trace to the appropriate limits in Table 12 on page 43. This can be done using Limit Lines (press Marker/Analysis > Analysis > Limit Test). The trace should be above the limit. PASS will be displayed on the screen if the limit lines are used. 11.Repeat step 9 and step 10 for Cal Sets “999.1” thru “999.12” (12-Port).
U3025AE10 Operational Check Troubleshooting Operational Check Failures If your test results fail the Operational Check limits, verify following: 1. Insure that the test set is turned on and connected properly to the PNA. 2. Check all appropriate network analyzer and test set connectors for damage, cleanliness, and proper torque. 3. Repeat the relevant 1-Port calibrations. 4. Verify that the stand-alone network analyzer is operating properly and meeting its published specifications.
U3025AE10 Theory of Operation Theory of Operation The following is a description of the operation of the U3025AE10. Reference the U3025AE10 block diagrams shown in Figure 55 on page 58. This section assumes the user has a general understanding of couplers, switches, and network analyzers. RF Switch Components All RF switches are mechanical. The switches select the RF paths from the PNA Source and Receiver through interconnect cables to Test Set port paths 3 through 12.
U3025AE10 Theory of Operation System Block Diagrams Figure 55 58 Standard Configuration User’s Guide
U3025AE10 Service Information Service Information Refer to “Contacting Agilent” on page 65. WARNING No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock do not remove covers. WARNING These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.
U3025AE10 Safety and Regulatory Information Safety and Regulatory Information Introduction Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition. Before Applying Power Verify that the product is configured to match the available main power source.
U3025AE10 Safety and Regulatory Information Compliance with German Noise Requirements This is to declare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordnung-3. GSGV Deutschland). Acoustic Noise Emission/Geraeuschemission LpA<70 dB Lpa<70 dB Operator Position am Arbeitsplatz Normal Operation normaler Betrieb per ISO 7779 nach DIN 45635 t.
U3025AE10 Safety and Regulatory Information Warnings WARNING The WARNING notice denotes a hazard. It calls attention to a procedure, practice, or the like, which if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. Warnings applicable to this instrument are: WARNING To prevent electrical shock, disconnect the Agilent Technologies U3025AE10 from the mains before cleaning.
U3025AE10 Safety and Regulatory Information Cautions CAUTION The CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like, which if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. Cautions applicable to this instrument are: CAUTION Always use the three-prong ac power cord supplied with this instrument.
U3025AE10 Safety and Regulatory Information Instrument Markings The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation. This symbol indicates that the instrument requires alternating current (ac) input. This symbol indicates separate collection for electrical and electronic equipment, mandated under EU law as of August 13, 2005.
U3025AE10 Contacting Agilent Contacting Agilent Assistance with test and measurement needs, and information on finding a local Agilent office are available on the Internet at: http://www.agilent.com/find/assist You can also purchase accessories or documentation items on the Internet at: http://www.agilent.com/find If you do not have access to the Internet, contact your field engineer.
U3025AE10 66 Contacting Agilent User’s Guide