Technical data

3-24 Service Guide N5230-90025
Tests and Adjustments PNA Series Microwave Network Analyzers
System Verification N5230C
Verification Kit box. Refer to Figure 3-9.
5. Under Printer Output, click one of the following options. Refer to Figure 3-9.
Print Tabular Data: Prints the verification data in tabular form which includes measured data and
uncertainty limits. For an example, refer to Figure 3-11 on page 3-26.
Print Graphs: Prints the verification data in graphical form. The graphical form includes the measured
data trace, factory supplied data trace, and uncertainty limits. For an example, refer to Figure 3-12 on
page 3-27.
File Tabular Data: Writes the tabular data to a text file in the C:\Program Files\Agilent\Network
Analyzer\Documents\ directory.
File Graphs: Saves a screen image in PNG format in the C:\Program Files\Agilent\Network
Analyzer\Documents\ directory.
NOTE For printed output, it is assumed that the printer has been tested and the Windows driver is
installed for the printer that is being used. The system verification test prints to the printer
that has been designated as the default printer. (On the Windows Desktop display, click on
My Computer, Control Panel, and then Printers to verify the printer setup.)
To modify the number of ports to be verified or to change the number of devices to measure, click on
the Configure tab and make the desired selections.
6. Click Run.
7. Follow the instructions on the analyzer for performing a full calibration or recalling an existing recent
calibration.
8. Follow the instructions on the analyzer for performing the system verification, inserting the verification
devices as prompted.