Technical data
Service Guide N5230-90025 2-5
PNA Series Microwave Network Analyzers General Product Information
N5230C Analyzer Options Available
Option 1E1, Source Attenuator (Extended Power Range)
This option is included in Options 145, 146, 245, and 246.
This option adds a single 60-dB step attenuator in the signal path of the measurement ports. This step
attenuator is used to adjust the power level (in 10 db steps) to the device under test (DUT) without changing
the power in the reference path.
Option 897, Built-In Tests for Commercial Calibration
This option enables performance tests that are built into the analyzer that enable you to perform ISO 9001
commercial calibration of your analyzer. Refer to “Service and Support Options” on page 2-10 for a
description of the commercial calibration.
Option 898 Built-In Tests for Standards Compliant Calibration
This option enables performance tests that are built into the analyzer that enable you to perform ISO 17025
and ANSI Z540 standards compliant calibration of your analyzer. Refer to “Service and Support Options” on
page 2-10 for a description of the standards compliant calibrations.
Option F13P04, 13.5 GHz Frequency Range
This is a 4-port network analyzer with an upper frequency limit of 13.5 GHz. This option is included in
Options 140, 145, and 146.
Option F20P04, 20 GHz Frequency Range
This is a 4-port network analyzer with an upper frequency limit of 20 GHz. This option is included in Options
240, 245, and 246.
Option S02, Second Source
This options adds a second source output with an internal second 60-dB step attenuator in this second
source signal path. This option is included in Options 146 and 246.