User`s guide
User’s Guide U3025-90001 37
U3025AE10 Operational Check Procedures
10.Compare the Reflection Tracking (1,1) trace to the appropriate limits in Table 9 on
page 32. This can be done using Limit Lines, select Marker/Analysis > Analysis > Limit
Test > Limit Test ON and Limit Line ON > Global pass/fail display ON, select Show
Table . The trace should be above the limit. PASS will be displayed on the screen if the
limit lines are used.
11.Repeat step 9 and step 10 for Cal Sets “999.1” thru “999.12” (12-Port).
NOTE For Ports 3-12 the response from 10 MHz to 500 MHz is normal due to the
PNA-X Couplers in comparison to the U3022AE10 bridges. The bridges have
more gain in the coupled RF path.