User`s guide
8 User’s Guide U3025-90001
U3025AE10 General Specifications
Frequency Range and Maximum Power Levels
The U3025AE10 frequency range is 10 MHz to 50 GHz.
CAUTION It is recommend that you do not operate components near damage levels
(+30 dBm). The power levels must be 3 dB below maximum level to ensure no
damage, see Table 3.
NOTE Refer to your PNA specifications to determine the maximum input power
levels for the access and test ports, or to optimize the power levels in the
receivers.
NOTE Damage and maximum levels are not necessarily the optimum level.
The examples shown assumes a 10 dBm (0.01w) RF signal and 7 VDC.
Equation 1 DC Wattage (
E
2
/R = DC Wattage)
Equation 2 Maximum Wattage (RF + DC = Maximum Wattage)
Table 3 Power Levels
Test Setup Power Level
Maximum U3025AE10 Multiport Test Set RF Power Levels:
PORT 3-12 +30 dBm 40 Vdc
SOURCE OUT +30 dBm 0 Vdc
CPLR ARM +30 dbm 0 Vdc
CPLR THRU +30 dBm 0 Vdc
RVCR A IN and B IN
1
1. Refer to your PNA specifications to determine the maximum input
power levels for the access and test ports.
+15 dBm 0 Vdc
Damage Power Levels to U3025AE10 Access and Test Ports:
2
2. Add the RF signal to the DC component to determine the maximum
wattage.
Max Level to Port 1 & 2 Test Ports +30 dBm 40 Vdc
7v
2
50⁄Ω0.98w=
0.01w 0.98w+ 0.99w=