U3042AE12 Operational Check
Preparing the N5242A or N5230C
1. Connect the Test Set to the N5242A 4-Port PNA-X using the interconnect cables as
shown in Figure 29 on page 39 and Table 11 on page 38. If you are using a N5230C,
refer to Figure 50 on page 69 and Table 10 on page 16.
2. Turn On the Test Set.
3. On the PNA, press Response > CAL > Manage CALS > CAL Set. Delete or Rename any
Cal Sets titled “999.1” thru “999.16” (16-Port), although it is unlikely that you will find
Cal Sets with these names.
4. Verify that the PNA is in 16-Port. See the bottom of the measurement window.
a. If only four S-Parameters are listed, press Utility > System > Configure > Multiport
Capability. On the Multiport Restart dialog, select Restart as multiport PNA with
this test set. Select U3042AE12 (16-Port).
5. Press Preset.
6. Verify that the [Stop Frequency] is set to [26.5 GHz], or [20 GHz] for the N5230C.
7. Verify that the [Start Frequency] is set to [10 MHz].
8. Verify that the [Power] is to set to [5 dBm].
9. Select Response > Avg > IF Bandwidth > 100 Hz. See Figure 58.
10.Select Stimulus > Sweep > Number of Points > 401.
11.Connect the ECal module to the PNA USB port on the front or rear panel, refer to
Figure 50 on page 69. This procedure assumes you are using a ECal. If you are not, see
“N5242A or N5230C 1-Port Calibration and Verification Procedure” step 2.
12.Allow the ECal module, Test Set and PNA to warm up for a minimum of 30 minutes.
Figure 58 Setting the IF Bandwidth