PRBS Generator 10 Gb/s User Guide
4.1.1 Performance Recommendations
Follow the following recommendations for best performance:
1. When using differential mode connection for OUT/OUT¯¯¯, ensure the
cables are phase balanced. If the electrical length of one cable is a
significant fraction of a unit interval longer than the other, the quality
of the differential signal will be degraded.
2. Keep cable lengths short and minimize number of cable bends.
3. When using a single port of differential output channel for single-
ended measurements, the complementary port must be terminated
with a 50 Ω termination.
4.1.2 Connector Care
The N4970A features high-quality SMA connectors for the front and rear
panel Input and Output, RF connections. Connector damage will degrade
signal fidelity.
Agilent Technologies also recommends the following:
• Use a 7 to 10 in-lbs torque wrench when attaching connectors.
•
Consider using connector savers to prolong performance and minimize
damage.
• Differential connectors may be used single-ended if second end
terminated in 50 Ω.
Inspect the connectors for the following:
• Worn or damaged threads
• Scratches to mating surface
• Burrs and loose metal particles
Clean the connectors as described in the following procedure. Cleaning
connectors with alcohol shall only be done with the instruments power cord
removed, and in a well-ventilated area. Allow all residual alcohol moisture to
evaporate, and the fumes to dissipate prior to energizing the instrument.