PRBS Generator 10 Gb/s User Guide
4.1.1 Performance Recommendations
Follow the following recommendations for best performance:
1. When using differential mode connection for OUT/OUT¯¯¯, ensure the
cables are phase balanced. If the electrical length of one cable is a
significant fraction of a unit interval longer than the other, the quality
of the differential signal will be degraded.
2. Keep cable lengths short and minimize number of cable bends.
3. When using a single port of differential output channel for single-
ended measurements, the complementary port must be terminated
with a 50 termination.
4.1.2 Connector Care
The N4970A features high-quality SMA connectors for the front and rear
panel Input and Output, RF connections. Connector damage will degrade
signal fidelity.
Agilent Technologies also recommends the following:
Use a 7 to 10 in-lbs torque wrench when attaching connectors.
Consider using connector savers to prolong performance and minimize
damage.
Differential connectors may be used single-ended if second end
terminated in 50 .
Inspect the connectors for the following:
Worn or damaged threads
Scratches to mating surface
Burrs and loose metal particles
Clean the connectors as described in the following procedure. Cleaning
connectors with alcohol shall only be done with the instruments power cord
removed, and in a well-ventilated area. Allow all residual alcohol moisture to
evaporate, and the fumes to dissipate prior to energizing the instrument.