Specifications

Making Your First Measurements
Rack Mounting the Test Set
Chapter 2 39
Rack Mounting the Test Set
Disconnect Device Requirements
In rack mounting operation a disconnect device, for example, an IEC320 appliance
coupler, must be used. Consideration must be given to the following when installing
the Test Set in a rack:
An adequately rated switch (250V, 10A) or circuit breaker (250V, 5A) shall be
included in the rack.
It should in close proximity to the Test Set and within easy reach of the operator.
It shall be marked as a disconnect device.
Ventilation Requirements
When installing the Test Set in a cabinet, the convection into and out of the Test Set
must not be restricted. If the total power dissipated in the cabinet is greater than 800
Watts, then forced convection must be used.
Rack Mounting
Using the Option N4010A-AX4 rack mount kit
1. Remove the carrying handle.
2. Remove the front and rear rubber bumpers.