Specifications
Making Your First Measurements
Rack Mounting the Test Set
Chapter 2 39
Rack Mounting the Test Set
Disconnect Device Requirements
In rack mounting operation a disconnect device, for example, an
IEC320 appliance coupler, must be used. Consideration must be given
to the following when installing the Test Set in a rack:
• An adequately rated switch (250V, 10A) or circuit breaker (250V,
5A) shall be included in the rack.
• It should in close proximity to the Test Set and within easy reach
of the operator.
• It shall be marked as a disconnect device.
Ventilation Requirements
When installing the Test Set in a cabinet, the convection into and out
of the Test Set must not be restricted. If the total power dissipated in
the cabinet is greater than 800 Watts, then forced convection must be
used.
Rack Mounting
Using the Option N4010A- AX4 rack mount kit
1. Remove the carrying handle.
2. Remove the front and rear rubber bumpers.