Technical data
44 N2780/81/82/83 User’s and Service Guide
• Amplitude:
• Offset: 0.0 Vdc
6 Measure the voltage output of the probe on the DMM and
record this value in Table 2.
7 Do the current calculation by multiplying the DMM
Voltage by the value in the Multiplier column of Table 2
and record the answer in the Calculated Amperes column
for the probe you are measuring.
8 Compare this value to the Passing Range values to
determine if the probe passes the test.
Table 2
N2780A/B N2781A/B N2782A/B N2783A/B
6.5 A rms 3 A rms 6 A rms 6 A rms
Model Number DMM Voltage Multiplier Calculated Amperes Passing Range
N2780A/B 100 505.5 A to 494.5 A
N2781A/B 100 151.6 A to 148.4 A
N2782A/B 10 30.31 A to 29.69 A
N2783A/B 10 30.31 A to 29.69 A
medium_standard.book Page 44 Friday, September 10, 2010 2:45 PM