User`s guide

70 Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
4 Math Waveforms
You can make peak-to- peak, maximum, minimum, and average dB
measurements on the FFT waveform. You can also find the frequency
value at the first occurrence of the waveform maximum by using the X
at Max Y measurement.
The following FFT spectrum was obtained by connecting a 4 V, 75 kHz
square wave to channel 1. Set the horizontal scale to 50 µs/div, vertical
sensitivity to 1 V/div, Units/div to 20 dBV, Offset to - 60.0 dBV, Center
frequency to 250 kHz, frequency Span to 500 kHz, and window to
Hanning.
See Also "To perform a transform function on an arithmetic operation" on
page 64
"FFT Measurement Hints" on page 71
"FFT Units" on page 72
"FFT DC Value" on page 72
"FFT Aliasing" on page 72
"FFT Spectral Leakage" on page 74
"Units for Math Waveforms" on page 74