236 Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
18 Utility Settings
4 Connect a BNC cable (40 inches maximum) from the TRIG OUT
connector on the rear panel to the BNC barrel connector.
5 Press the [Utility] key; then, press the Service softkey.
6 Begin the Self Cal by pressing the Start User Cal softkey.
To perform hardware self test
Pressing [Utility] > Service > Hardware Self Test performs a series of internal
procedures to verify that the oscilloscope is operating properly.
It is recommended you run Hardware Self Test:
After experiencing abnormal operation.
For additional information to better describe an oscilloscope failure.
To verify proper operation after the oscilloscope has been repaired.
Successfully passing Hardware Self Test does not guarantee 100% of the
oscilloscope's functionality. Hardware Self Test is designed to provide an
80% confidence level that the oscilloscope is operating properly.
To perform front panel self test
Pressing [Utility] > Service > Front Panel Self Test lets you test the front panel
keys and knobs as well as the oscilloscope display.
Follow the on- screen instructions.
To display oscilloscope information
Press [Help] > About Oscilloscope to display information about your
oscilloscope:
Model number.
Serial number.
Bandwidth.
Module installed.
Software version.
Installed licenses. See also "Loading Licenses and Displaying License
Information" on page 261.