User`s guide

Utility Settings 18
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide 235
For a 4- channel oscilloscope, connect BNC tees to the equal- length
cables as shown below. Then connect a BNC(f)- to- BNC(f) (barrel
connector) to the tee as shown below.
Figure 36 User Calibration cable for 2-channel oscilloscope
Figure 37 User Calibration cable for 4-channel oscilloscope
To Channel 1
To Channel 2
Long cable
to TRIG OUT
To Channel 4
To Channel 1
To Channel 3
Longer cable
to TRIG OUT
T
o Channel 2