16 Save/Recall (Setups, Screens, Data)
4 If you would like to clear the display, press Clear Display.
To recall mask files
Mask files can be recalled from one of four internal (\Agilent Flash)
locations or from an external USB storage device.
1 Press [Save/Recall] > Recall > Recall:; then, turn the Entry knob to select
Mask (*.msk).
2 Press the softkey in the second position and use the Entry knob to
navigate to the file to recall. See "To navigate storage locations" on
page 214.
3 Press the Press to Recall softkey.
A message indicating whether the recall was successful is displayed.
4 If you would like to clear the display or clear the recalled mask, press
Clear Display or Clear Mask.
To recall reference waveform files from a USB storage device
1 Press the [Save/Recall] key.
2 In the Save/Recall Menu, press the Recall softkey.
3 In the Recall Menu, press the Recall softkey and turn the Entry knob to
select Reference Waveform data (*.h5) .
4 Press the To Ref: softkey and turn the Entry knob to select the desired
reference waveform location.
5 Press the softkey in the second position and use the Entry knob to
navigate to the file to recall. See "To navigate storage locations" on
page 214.
6 Press the Press to Recall softkey.
A message indicating whether the recall was successful is displayed.
7 If you would like to clear the display of everything except the reference
waveform, press Clear Display.
Recalling Default Setups
1 Press the [Save/Recall] key.