Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide 179
Time Measurements
The following figure shows time measurement points.
The default lower, middle, and upper measurement thresholds are 10%,
50%, and 90% between Top and Base values. See "Measurement
Thresholds" on page 184 for other percentage threshold and absolute value
threshold settings.
"Period" on page 179
"Frequency" on page 180
"+ Width" on page 181
"– Width" on page 181
"Burst Width" on page 181
"Duty Cycle" on page 181
"Rise Time" on page 182
"Fall Time" on page 182
"Delay" on page 182
"Phase" on page 183
Period
Period is the time period of the complete waveform cycle. The time is
measured between the middle threshold points of two consecutive,
like-polarity edges. A middle threshold crossing must also travel through
the lower and upper threshold levels which eliminates runt pulses. The X
cursors show what portion of the waveform is being measured. The Y
cursor shows the middle threshold point.
Rise Time
+ Width
Fall Time
- Width
Period
Middle
Upper
Thresholds
Lower