Agilent InfiniiVision 2000 X-Series Oscilloscopes
User's Guide
10
Trigger Mode/Coupling
To select the Auto or Normal trigger mode 134
To select the trigger coupling 136
To enable or disable trigger noise rejection 137
To enable or disable trigger HF Reject 137
To set the trigger holdoff 138
External Trigger Input 138
To access the Trigger Mode and Coupling Menu:
• In the Trigger section of the front panel, press the [Mode/Coupling] key.
Noisy Signals If the signal you are probing is noisy, you can set up the oscilloscope to
reduce the noise in the trigger path and on the displayed waveform. First,
stabilize the displayed waveform by removing the noise from the trigger
path. Second, reduce the noise on the displayed waveform.
1 Connect a signal to the oscilloscope and obtain a stable display.
2 Remove the noise from the trigger path by turning on high- frequency
rejection ("To enable or disable trigger HF Reject" on page 137),
low- frequency rejection ("To select the trigger coupling" on page 136),
or "To enable or disable trigger noise rejection" on page 137.
3 Use "Averaging Acquisition Mode" on page 151 to reduce noise on the
displayed waveform.