User`s guide
15
Calibration Test Record
Calibration Test Record
Agilent Technologies
N2790A 100 MHz Differential Probe
Serial No.:_______________________
Certification Date:_________________
Tested By:_______________________
Recommended Test Interval: 1 Year
Recommended Date of Next Certification:_________
Certification Temperature:_____________________
_______________________________
_______________________________
Test Probe Settings Limits Results
Gain 50:1 +196 mV ... +204 mV (\2% of reading)
Gain 50:1 +1.96 V ... +2.04 V (\2% of reading)
Gain 500:1 +196 mV ... +204 mV (\2% of reading)
Gain 500:1 +1.96 V ... +2.04 V (\2% of reading)
Gain 50:1 -196 mV ... -204 mV (\2% of reading)
Gain 50:1 -1.96 V ... -2.04 V (\2% of reading)
Gain 500:1 -196 mV ... -204 mV (\2% of reading)
Gain 500:1 -1.96 V ... -2.04 V (\2% of reading)
Bandwidth 50:1 > 2.8 div Reading on oscilloscope
Bandwidth 500:1 > 2.8 div Reading on oscilloscope