User`s guide

13
Using the N2795A/N2796A Probes
Oscilloscope Compatibility
The N2795A/N2796A probes are compatibile with the following Agilent oscilloscopes
(those with the Agilent AutoProbe Interface):
InfiniiVision 5000 Series, software version 6.10 or higher
InfiniiVision 6000 Series (except 100 MHz model), software version 6.10 or higher
InfiniiVision 7000 Series, software version 6.10 or higher
Infiniium 9000 Series, software version 2.50 or higher
Infiniium 90000 Series, software version 2.50 or higher
Infiniium 90000 X-Series (requires N5442A adapter)
General Measurement Procedure / Functional Test
The following is a generic measurement procedure you can use to ensure your unit is
functioning properly.
1 Connect the probe to an oscilloscope channel input and ensure the input
impedance of the oscilloscope matches the output impedance of the probe
(50W).
2 Equip the probe with a grounding accessory and connect the ground to the
ground terminal on the oscilloscope.
3 Connect the probe tip to the Probe Compensation Output on the oscilloscope.
4 Enable autoscale on the oscilloscope.
5 A square wave should now be displayed on the oscilloscope if the probe is
functioning properly.