User`s guide

Agilent N5161A/62A/81A/82A/83A MXG Signal Generators User’s Guide 135
Using Pulse Modulation (Option UNU or UNW or 320)
Pulse Train (Option 320 – Requires: Option UNU or UNW)
Figure 6-6 Pulse Train: Import From Selected File Softkeys
Pulse > Pulse Source > More > Pulse
Train > Edit Pulse Train > More
page 65
These softkeys delete
individual On Time or Off
Time elements as well as the
Repeat cycle counts.
This file’s data structure uses decimals in the numbers
and so requires the Import Decimal Separator to be set to
Dot.
Selects whether the decimal point
is a "." or a "," during import of
CSV/ASCII files.
This value is persistent across
preset/recall and power cycles.
SCPI Commands:
:MEMory:IMPort[:ASCii]:SEParato
r:DECimal DOT|COMMa
:MEMory:IMPort[:ASCii]:SEParato
r:DECimal?
After confirming import of the file
the new file’s pulse train values
are displayed.
Deleting all Pulse Cycle rows
(elements) must be
confirmed.
For details on each key, use key help as described on page 42.
Note: Column
separated values are
auto-detected by the
MXG.
page 136
page 136