Technical data
33
Memory Clearing, Sanitization and Removal Procedures
Force Internal to SD Card
NOTE The Erase and Sanitize All operation resets the instrument’s LAN settings to their
factory defaults. For more details, see “LAN Setup” on page 34.
Force Internal to SD Card
This function is available only in the MXG Series B and EXG instruments.
The options for this setting are Off or On:
• On: Forces all internal settings and files to be stored only on the external SD memory card, including
instrument states, user data files, sweep list files, and other user created files. If it becomes necessary to
sanitize the instrument, these settings may then be securely retained by removing the SD card and
retaining it in the secure area.
• Off: All internal settings are stored in the instrument’s Main Memory (see “MXG (Series B) & EXG
Memory” on page 16). This memory may be sanitized using the Erase and Sanitize All function.
Model Key Sequence
MXG (Series A/B)
EXG
File > More > Security > Erase and Sanitize All > Confirm Sanitize
PSG
ESG
Utility > Memory Catalog > More > Security > Erase and Sanitize All >
Confirm Sanitize
Model Key Sequence
MXG (Series B)
EXG
File > More > Security > Force Internal to SD Card