System Installation and Verification
Installation and Verification Installation and Verification
Installation and Verification
28
2828
28
If the reading is not less than 20mv, allow more time for the Vacion power supply to operate.
Recheck this again before proceeding.
Failure to do this step could result in permanent damage to the X-ray tube. The X-ray tube has
components that need to warm up before generating X-rays.
8
88
8 From the operator control panel (Figure 22) turn the X-
Ray Enable Key to Enable, and then push the X-Ray On
Button.
9
99
9 Go to the Service environment of the Medalist x6000
software by selecting Window > Service.
10
1010
10 Select the Subsystem, Status and Control task.
11
1111
11 Select the X-ray Source tab and click Initialize. This
will set the X-ray voltage to 160kv and current to 0
micro Amps.
Once the high voltage has ramped to about 160 kv in
the status display continue with the procedure.
Current should be set to 0 micro amps at this point.
12
1212
12 Set the X-ray tube deflection coil voltages on the
E3620A for V1 to 12 Volts and V2 to 12 Volts. This is a
midrange value only used for the X-ray Safety Test.
13
1313
13 Set the focus coil current to the value from the X-ray
tube label. See Figure 24.
CAUTION
CAUTION
NOTE